Description
TCM stores calibration data (inherent characteristic value) of G sensor to provide accurate control. Therefore, it is required to perform calibration of G sensor after the following work is performed.
Work Procedure
1.PREPARATION BEFORE CALIBRATION PROCEDURE
>> GO TO 2.
2.PERFORM CALIBRATION
With CONSULT
CAUTION: Never start engine.
CAUTION: Never swing the vehicle during "G sensor calibration".
Is "COMPLETED" displayed?
YES >> GO TO 3.
NO >> Perform steps 1 and 2 again.
3.PERFORM THE SELF-DIAGNOSIS
With CONSULT
Is "P1586" or "P1588" detected?
YES >> Go to TM "DTC Index".
NO >> Calibration end

Cleaning Whenever an automatic transaxle is repaired, overhauled, or replaced, the CVT fluid cooler mounted in the radiator must be inspected and cleaned. Metal debris and friction material, i ...
EVAP control system pressure sensor
Exploded View
1. EVAP control system pressure sensor 2. O-ring 3. EVAP canister
Removal and Installation
NOTE:
The EVAP canister system pressure sensor can be removed without removing the
EVAP canister.
REMOVAL
Remove the EVAP canister protector cover.
Disconnect EVAP canister purg ...
Line pressure test
Work Procedure
INSPECTION
Inspect the amount of engine oil. Replenish the engine oil if necessary.
Refer to LU, "Inspection".
Drive for about 10 minutes to warm up the vehicle so that the A/T fluid
temperature is to 50 to 80C (122
to 176F).
Inspect the amount of ATF. Rep ...